1 - Intubate as usual with an endotracheal (ET) tube size 7.0 mm or larger. Attach the EZ-Multiport Adaptor.
2 - Confirm depth of intubation, leaving 4 cm between distal tip of the ET tube and the carina to allow sufficient space for the deployment of the bifurcated tips.
3 - Insert the EZ-Blocker™ Endobronchial Blocker in the EZ-Multiport™ Adaptor and then reinsert the bronchoscope. Advance the EZ-Blocker™ Endobronchial Blocker until the bifurcated ends are in either bronchus. Verify device orientation under visualization by inflating and deflating cuffs.
4 - Before the surgeon breaks the thoracic vacuum, ensure both cuffs of the EZ-Blocker™ Endobronchial Blocker are deflated and discontinue ventilation until lung collapse is achieved. Inflate the cuff in the bronchus of the lung to be isolated and then resume ventilation.